You are seeing this message because your Web browser does not support basic Web standards. Find out more about why this message is appearing and what you can do to make your experience on this site better.


ABOUT JAMA
Advanced Search

Welcome   | My Account | E-mail Alerts | Access Rights | Sign In


  Vol. 295 No. 16, April 26, 2006 TABLE OF CONTENTS
  JAMA
  •  Online Features
  Review
 This Article
 •Full text
 •PDF
 •Send to a friend
 • Save in My Folder
 •Save to citation manager
 •Permissions
 Citing Articles
 •Citation map
 •Citing articles on HighWire
 •Citing articles on Web of Science (25)
 •Contact me when this article is cited
 Related Content
 •Related articles
 •Similar articles in JAMA
 Topic Collections
 •Pacemakers/ Defibrillators
 •Cardiovascular System
 •Quality of Care
 •Evidence-Based Medicine
 •Quality of Care, Other
 •Review
 •Arrhythmias
 •Cardiovascular Intervention
 •Alert me on articles by topic
 Social Bookmarking
  Add to CiteULike Add to Connotea Add to Del.icio.us Add to Digg Add to Reddit Add to Technorati Add to Twitter What's this?

Pacemaker and ICD Generator Reliability

Meta-analysis of Device Registries

William H. Maisel, MD, MPH

JAMA. 2006;295:1929-1934.

Context  Despite there being millions of pacemaker and implantable cardioverter-defibrillator (ICD) generator implants worldwide, little is known about device reliability.

Objectives  To perform a meta-analysis of prospective pacemaker and ICD registries to determine annual rates of pacemaker and ICD malfunction and to identify trends in these rates.

Data Sources  MEDLINE (January 1966 to April 2005), the Cochrane Central Register of Controlled Trials (through second quarter 2005), the Cochrane Database of Systematic Reviews (through second quarter 2005), and a bibliographic review of secondary sources. Search terms included pacemaker, artificial; defibrillators, implantable; registries; performance; and malfunction.

Study Selection  Eligible registries were prospective; reported the number of patients with pacemaker, ICD, or both; and allowed determination of the annual number of device malfunctions. Of 1007 references screened, 3 registries meeting the selection criteria were identified and included 2.1 million pacemaker person-years and 14 821 ICD person-years of observation.

Data Extraction  Included the annual number of patients with pacemakers and ICDs at risk of device failure and the annual number of generator malfunctions (1983-2004 for pacemakers, 1988-2004 for ICDs). A device malfunction was defined as an integral component failure that required device explantation prior to reaching elective replacement. Failures of pacemaker and ICD electrodes were not included in the study.

Data Synthesis  There were 2981 pacemaker and 384 ICD generator malfunctions. Pacemaker reliability improved markedly during the 1980s (P for trend <.001) and the pacemaker malfunction rate remained low during the remainder of the study. Implantable cardioverter-defibrillator reliability improved during the first 10 study years (P for trend <.001). From 1998-2002, however, the ICD malfunction rate increased more than 4-fold (P for trend <.001), before decreasing substantially in the latter 2 years of the study. Overall, the mean (SE) annual ICD malfunction rate was about 20-fold higher than the pacemaker malfunction rate (26.5 [3.8] vs 1.3 [0.1] malfunctions per 1000 person-years, P<.001). Battery malfunctions were the most common cause of device failure.

Conclusions  Pacemaker reliability has improved markedly. In contrast, after more than a decade of improving device reliability, the ICD malfunction rate transiently increased before experiencing substantial reductions in the latter 2 study years. Whether increasing device sophistication accounts for the observed decrease in reliability is not known. Continued monitoring of pacemaker and ICD performance is required.


Author Affiliation: Cardiovascular Division, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Mass.



Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati   Add to Twitter Twitter     What's this?

RELATED ARTICLES

Pacemaker and ICD Generator Malfunctions: Analysis of Food and Drug Administration Annual Reports
William H. Maisel, Megan Moynahan, Bram D. Zuckerman, Thomas P. Gross, Oscar H. Tovar, Donna-Bea Tillman, and Daniel B. Schultz
JAMA. 2006;295(16):1901-1906.
ABSTRACT | FULL TEXT  

Complications Associated With Implantable Cardioverter-Defibrillator Replacement in Response to Device Advisories
Paul A. Gould, Andrew D. Krahn, and for the Canadian Heart Rhythm Society Working Group on Device Advisories
JAMA. 2006;295(16):1907-1911.
ABSTRACT | FULL TEXT  

Pacemaker and ICD Malfunction—An Incomplete Picture
Bruce L. Wilkoff
JAMA. 2006;295(16):1944-1946.
EXTRACT | FULL TEXT  

Implantable Cardioverter-Defibrillators
Janet M. Torpy, Alison E. Burke, and Richard M. Glass
JAMA. 2006;295(16):1964.
EXTRACT | FULL TEXT  


THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES

Transtelephonic Versus Remote Monitoring of Cardiovascular Implantable Electronic Devices Is One Approach to Be Preferred?
Schoenfeld
J Am Coll Cardiol 2009;54:2020-2022.
FULL TEXT  

Towards predictive modelling of the electrophysiology of the heart
Vigmond et al.
Exp Physiol 2009;94:563-577.
ABSTRACT | FULL TEXT  

Heart Disease and Stroke Statistics--2009 Update: A Report From the American Heart Association Statistics Committee and Stroke Statistics Subcommittee
WRITING GROUP MEMBERS et al.
Circulation 2009;119:e21-e181.
FULL TEXT  

Performance and survival of transvenous defibrillation leads: need for a European data registry
Goette et al.
Europace 2009;11:31-34.
ABSTRACT | FULL TEXT  

No long-term psychological morbidity living with an implantable cardioverter defibrillator under advisory: the Medtronic Marquis experience
Birnie et al.
Europace 2009;11:26-30.
ABSTRACT | FULL TEXT  

The efficacy of ventricular pacing with device automaticity in paediatric patients
Tomaske et al.
Europace 2008;10:838-843.
ABSTRACT | FULL TEXT  

Symptomatic heart failure is the most important clinical correlate of impaired quality of life, anxiety, and depression in implantable cardioverter-defibrillator patients: a single-centre, cross-sectional study in 610 patients
Johansen et al.
Europace 2008;10:545-551.
ABSTRACT | FULL TEXT  

Tunnel Propagation of Postshock Activations as a Hypothesis for Fibrillation Induction and Isoelectric Window
Ashihara et al.
Circ. Res. 2008;102:737-745.
ABSTRACT | FULL TEXT  

The Year in Epidemiology, Health Services Research, and Outcomes Research
Krumholz and Masoudi
J Am Coll Cardiol 2007;50:2254-2262.
FULL TEXT  

Sudden Death Prevention With Implantable Devices
Passman and Kadish
Circulation 2007;116:561-571.
FULL TEXT  

Transvenous Implantable Cardioverter-Defibrillator Leads: The Weakest Link
Maisel
Circulation 2007;115:2461-2463.
FULL TEXT  

Update in Critical Care 2006
Milbrandt et al.
Am. J. Respir. Crit. Care Med. 2007;175:638-648.
FULL TEXT  

Contemporary Pacemaker and Defibrillator Device Therapy: Challenges Confronting the General Cardiologist
Schoenfeld
Circulation 2007;115:638-653.
FULL TEXT  

ICD and Pacemaker Malfunctions
Journal Watch Cardiology 2006;2006:9-9.
FULL TEXT  

Management of recalled pacemakers and implantable cardioverter-defibrillators: a decision analysis model.
Amin et al.
JAMA 2006;296:412-420.
ABSTRACT | FULL TEXT  

ICD and Pacemaker Malfunctions
Journal Watch Cardiology 2006;2006:2-2.
FULL TEXT  

Pacemaker and ICD generator malfunctions: analysis of Food and Drug Administration annual reports.
Maisel et al.
JAMA 2006;295:1901-1906.
ABSTRACT | FULL TEXT  

Pacemaker and ICD malfunction--an incomplete picture.
Wilkoff
JAMA 2006;295:1944-1946.
FULL TEXT  





HOME | CURRENT ISSUE | PAST ISSUES | TOPIC COLLECTIONS | CME | SUBMIT | SUBSCRIBE | HELP
CONDITIONS OF USE | PRIVACY POLICY | CONTACT US | SITE MAP
 
© 2006 American Medical Association. All Rights Reserved.