Manufacturer's Recall of Rapid Assay Kits Based on False Positive Cryptosporidium Antigen Tests Wisconsin, 2001-2002
JAMA. 2002;287:1798.
 |
 |
| Since this article does not have an abstract, we have provided the first 150 words of the full text and any section headings. |
|
 |
 |
MMWR. 2002;51:189
The Wisconsin Division of Public Health and the Wisconsin State Laboratory of Hygiene (WSLH) reported that a recent cluster of cryptosporidiosis cases in a three-county area in southeastern Wisconsin was the result of false-positive tests. During December 1, 2001February 1, 2002, approximately 30 cases of cryptosporidiosis were diagnosed at a laboratory in southeastern Wisconsin using the Becton, Dickinson, and Company (Franklin Lakes, New Jersey) ColorPACTM Cryptosporidium/Giardia rapid assay (lot number 219370, expiration date 2002-06-05). Seventeen stool specimens, which were collected from 11 patients and tested positive by the rapid assay, were re-evaluated at WSLH. Six of these stool specimens were in EcoFix (Meridian Bioscience Inc., Cincinnati, Ohio), eight were in Cary-Blair transport media, and three were formalin fixed. All 17 specimens tested negative for Cryptosporidium at WSLH using the hot safranin stain and MeriFluor (Meridian Bioscience Inc., Cincinnati, Ohio) Cryptosporidium/Giardia direct fluorescent antibody kit with concentrated specimens.
For . . . [Full Text of this Article]