Radiation Doses Associated With Cardiac Computed Tomography Angiography
 |
 |
| Since this article does not have an abstract, we have provided the first 150 words of the full text and any section headings. |
|
 |
 |
To the Editor: Dr Hausleiter and colleagues1 compared radiation doses resulting from different CCTA studies and concluded that identification of factors that influence radiation dose may be important. While technical characteristics clearly influence the radiation dose, major differences in radiation dose could also arise from the calculation methodology. For example, the authors calculated the effective dose using a region-specific factor of 0.014, whereas a study by Husmann et al2 used a conversion factor of 0.017. Using different methods for calculating the effective dose can be misleading. The flaw in the concept of effective dose calculation is mainly due to the disparity of the tissue-weighting factors, which stand at the foundation of effective dose calculation.3-5 To eliminate this discrepancy, DLP values rather than effective dose for different CT scenarios should be compared, because DLP calculation is standardized and it is currently displayed by a majority of the CT scanners. Moreover, the . . . [Full Text of this Article]
Razvan C. Opreanu, MD;
John P. Kepros, MD, MBA
john.kepros@hc.msu.edu Department of Surgery Michigan State University East Lansing
CiteULike Connotea Del.icio.us Digg Reddit Technorati Twitter
What's this?
RELATED ARTICLE
Estimated Radiation Dose Associated With Cardiac CT Angiography
Jörg Hausleiter, Tanja Meyer, Franziska Hermann, Martin Hadamitzky, Markus Krebs, Thomas C. Gerber, Cynthia McCollough, Stefan Martinoff, Adnan Kastrati, Albert Schömig, and Stephan Achenbach
JAMA. 2009;301(5):500-507.
ABSTRACT
| FULL TEXT
RELATED LETTERS
Radiation Doses Associated With Cardiac Computed Tomography Angiography
Joseph M. Price
JAMA. 2009;301(22):2324.
EXTRACT
| FULL TEXT
Radiation Doses Associated With Cardiac Computed Tomography Angiography
Iyad N. Daher
JAMA. 2009;301(22):2324.
EXTRACT
| FULL TEXT
Radiation Doses Associated With Cardiac Computed Tomography Angiography—Reply
Jörg Hausleiter, Thomas C. Gerber, and Stephan Achenbach
JAMA. 2009;301(22):2325.