Improving Survival From Sudden Cardiac Arrest
The Role of the Automated External Defibrillator
John P. Marenco, MD;
Paul J. Wang, MD;
Mark S. Link, MD;
Munther K. Homoud, MD;
N. A. Mark Estes III, MD
JAMA. 2001;285:1193-1200.
ABSTRACT
 |  |
Context Sudden cardiac death is a major public health problem in the United States, and improving survival after out-of-hospital cardiac arrest has been the subject of intense study. Early defibrillation has been shown to be critical to improving survival. Use of automated external defibrillators (AEDs) has become an important component of emergency medical systems, and recent advances in AED technology have allowed expansion of AED use to nontraditional first responders and the lay public.
Objectives To examine advancements in AED technology, review the impact of AEDs on time to defibrillation and survival, and explore the future role of AEDs in the effort to improve survival following sudden cardiac arrest.
Data Sources MEDLINE was searched for articles from 1966 through December 2000 (Medical Subject Headings: electric countershock, heart arrest, resuscitation, emergency medical services; keywords: automatic external defibrillator, automated external defibrillator, public access defibrillation). Reference lists of relevant articles, news releases, and product information from manufacturers were also reviewed.
Study Selection Initial MEDLINE search produced 4816 articles, from which 101 articles were selected for referencing based on having been published in a peer-reviewed journal and on relevance to the subject of the manuscript as determined by all 5 authors.
Data Extraction All studies were critically reviewed for relevance, accuracy, and quality of data and study design by all authors.
Data Synthesis Recent advances in AED technology and design have resulted in marked simplification of AED operation, improvements in accuracy and effectiveness, and reductions in cost. Use of AEDs by first responders and laypersons has reduced time to defibrillation and improved survival from sudden cardiac arrest in several communities. Initial studies of the cost-effectiveness of AED use in comparison with other commonly used treatments are favorable.
Conclusion The AED represents an efficient method of delivering defibrillation to persons experiencing out-of-hospital cardiac arrest and its use by both traditional and nontraditional first responders appears to be safe and effective. The rapidly expanding role of AEDs in traditional emergency medical systems is supported by the literature, and initial studies of public access to defibrillation offer hope that further improvements in survival after sudden cardiac death can be achieved.
INTRODUCTION
Cardiovascular disease is the major cause of death in the United States, resulting in nearly 1 million deaths a year. Nearly half of these deaths (250 000 to 500 000) are sudden and unexpected.1-3 Most sudden deaths from cardiac arrest occur outside the hospital, and survival rates have traditionally been pooronly 1% to 5% of these patients are estimated to survive to hospital discharge.4-5 When first-responders arrive early, ventricular tachycardia and ventricular fibrillation are the rhythms they most commonly encounter initially.1-2,7 Most persons experiencing cardiac arrest have no history of severe heart disease, and sudden cardiac death is frequently the first manifestation of cardiovascular disease (Figure 1).1, 6-7 This has complicated efforts at prevention and has led to a shift in emphasis to improvements in prehospital care.
|
|
|
|
Figure 1. Incidence of Sudden Death From Cardiac Arrest
A, Sudden death incidence and total events for various population pools, showing inverse relationship between risk and total number of events. Note that the horizontal axis for the incidence graph is not linear. B, Comparison of sudden death rates in a high-risk subgroup of patients with cardiovascular disease after a major cardiovascular event with a low-risk subgroup of patients with cardiovascular disease but free of a major event. After 18 months, the curves become parallel indicating that the presence of a prior major cardiovascular event as a risk factor for future cardiac arrest becomes limited with time. EF indicates ejection fraction; VT, ventricular tachycardia; VF, ventricular fibrillation; MI, myocardial infarction; and CV, cardiovascular. Reproduced with permission from Myerburg, et al.6
|
|
|
In an effort to improve prehospital care, the American Heart Association (AHA) has promoted the "Chain of Survival" concept, describing a sequence of interventions (links) that when implemented result in improved survival following sudden cardiac arrest.8 Early defibrillation has emerged as the single most important intervention.7, 9-11 There are data from both animal and human studies showing that defibrillation immediately after witnessed ventricular fibrillation results in survival rates greater than 90%.12-15 Each minute of ventricular fibrillation, however, leads to a nearly 10% reduction in survival. Although cardiopulmonary resuscitation (CPR) prior to defibrillation in prolonged arrests (>4 minutes) may improve survival,16 chances of long-term survival of patients defibrillated after 10 minutes are dismal (Figure 2).7, 9-11 Early use of external defibrillation by emergency medical technicians (EMTs) in selected communities with rapid response times has been shown to improve survival.17-18 However, many communities, both rural and urban, continue to have poor survival rates, presumably due to longer response times of emergency personnel. Studies of persons experiencing sudden cardiac arrest in New York City and Chicago, where emergency medical system (EMS) response times are prolonged by heavy traffic and tall buildings, have demonstrated survival rates of 2% or less.5, 19
|
|
|
|
Figure 2. Rates of Survival to Hospital Discharge
Rates given for 4 years of consecutive patients (1112) initially discovered in ventricular fibrillation. Response times are known for 942 cases. Both witnessed and unwitnessed cases in this 4-year period are considered. The average (SD) response time for first emergency unit was 3.0 (1.5) minutes and for paramedics was 6.5 (3.2) minutes after dispatch. VF indicates ventricular fibrillation. Reproduced with permission.9
|
|
|
In an effort to overcome these limitations, the AHA has promoted the concept of public access defibrillation (PAD).20-21 This concept promotes the expansion of the role of defibrillation to both minimally trained first-responders (police officers, firefighters, security guards, flight attendants) and to trained laypersons who witness an arrest. It also promotes the placement of automated external defibrillators (AEDs) in such areas as airports, convention centers, sporting arenas, casinos, shopping malls, and large office buildings. Some have envisioned a future where the AED is as commonplace as the fire extinguisher.2 One of the keys to making PAD feasible has been advances in technology over the past 2 decades that have made AED use by nonmedical personnel safe and effective. Although there is growing literature to suggest that PAD is greatly improving survival from sudden cardiac arrest, many questions remain. We will examine the important advances in AED technology over the past 2 decades, review the existing literature on the effectiveness of the AED, and explore the future role of the AED in our effort to improve survival from sudden cardiac arrest.
METHODS
We used MEDLINE to identify all English-language publications on AEDs from 1966 to December 2000. The medical subject headings used were electric countershock, heart arrest, resuscitation, and emergency medical services. Automatic external defibrillator, automated external defibrillator, and public access defibrillation were searched as keywords. All relevant publications were reviewed. Data quality was determined by publication in peer-reviewed literature. In addition, all 5 authors reviewed relevant abstracts and presentations from the official 1999 and 2000 annual meetings of the American College of Cardiology, the AHA, and the North American Society of Pacing and Electrophysiology. Product information was obtained from each AED manufacturer's official Web site and official specification sheets were provided by each of the manufacturers at our request.
RESULTS
Technology
Automated external defibrillators were developed in the 1970s and first introduced for clinical use in 1979.22-23 The AED is a lightweight portable device containing a battery, capacitors, and circuitry designed to analyze cardiac rhythm and inform the operator whether a shock is indicated. Information is transmitted to the device by electrode pads used for both monitoring and shock therapy (Figure 3). Manufacturers were challenged to develop an AED so reliable and easy to use that fears of misuse and inappropriate shocks would be unfounded. Reductions in size, weight, cost, and maintenance were also essential if PAD were to be logistically and economically feasible. Several advances in AED technology over the past decade have been instrumental in the effort to achieve these goals.
|
|
|
|
Figure 3. Automated External Defibrillator With Attached Electrode Pads
AED indicates automated external defibrillator; ECG, electrocardiogram.
|
|
|
Ease of Use
Laypersons trained to use an AED may go months or years without witnessing an arrest or operating an AED. Operation of an AED, therefore, needs to be nearly intuitive for timely delivery of therapy. Several important changes have resulted in marked simplification of AED use. Self-adhesive electrode pads are provided with diagrams on how to apply them (Figure 3). Once activated, AEDs have voice and text prompts to guide the user through the few simple steps. An arrhythmia analysis algorithm automatically interprets the rhythm and either recommends countershock, to be given by the push of a button, or no countershock. The device immediately reevaluates the rhythm and determines whether to recommend an additional shock. Cardiac rhythms are automatically recorded for review. These simplifications in AED operation have resulted in a marked reduction in defibrillation times and have minimized the need for retraining. A study examining the use of the AED in out-of-hospital cardiac arrests showed that trained first responders had an average time from power-on to first defibrillation of only 25 seconds.24 One recent study of mock cardiac arrest showed that mean time to defibrillation from arrival at the scene was only 90 seconds for a group of untrained sixth-grade students and 67 seconds for trained EMTs and paramedics.25
Arrhythmia Analysis Algorithms
Taking advantage of innovations in computer technology and detection algorithm design in the 1980s and 1990s, manufacturers have developed arrhythmia analysis algorithms that can interpret complex cardiac rhythms and deliver appropriate therapy with impressive accuracy (Figure 4). In 1997, the AHA Subcommittee on AED Safety and Efficacy recommended specific performance goals for arrhythmia analysis algorithms. Current AEDs have consistently exceeded these goals.23 Several studies have demonstrated 100% sensitivity and specificity for the detection of ventricular fibrillation.24, 26-29
|
|
|
|
Figure 4. Electrocardiograms From a Person in Cardiac Arrest
Ventricular fibrillation is detected (note the horizontal bars measure maximum amplitude in millivolts), charging begins (from first to third arrow), a shock is advised (second arrow), and a charge delivered (fourth arrow). An organized rhythm is detected and no further shocks advised.
|
|
|
Energy Delivery and Storage
External defibrillation requires the delivery of energy, in the form of current, to the myocardium. This process has been made more efficient through the use of impedance-based defibrillation, larger electrode pad sizes, and biphasic waveforms. Impedance-based defibrillation refers to the adjustment of shock waveform features or shock energy based on patient impedance (resistance). Because defibrillation thresholds vary substantially from patient to patient, this feature results in a more efficient use of energy.30-33 Larger electrode pad sizes have been shown to reduce transthoracic impedance and improve defibrillation success rates.34-35
Early AEDs and most external defibrillators used monophasic waveforms, in which current is delivered to the patient in a single direction (polarity). Two conventional monophasic waveforms exist: damped sinusoidal, in which a high peak current is delivered with the current returning to zero gradually; and truncated exponential, in which current returns to zero instantaneously after delivery of the selected energy. More recently, biphasic waveforms, in which the direction of current flow is reversed part way through the pulse (Figure 5), have been used extensively in implantable cardiac defibrillators (ICDs) and found to achieve equivalent or superior defibrillation rates at relatively lower energy levels (<200 J) than the previously used monophasic waveforms.36-48 While direct comparison of biphasic with monophasic waveforms in the out-of-hospital setting are lacking, evidence that lower energy shocks using biphasic waveforms have comparative efficacy and are safe and clinically effective has led some AED manufacturers to use a fixed 150-J shock energy.23, 30, 48 Poole et al29 demonstrated a first-shock defibrillation rate of 89% using low-energy (150 J) biphasic waveforms in patients with out-of-hospital cardiac arrest found to be in ventricular fibrillation, while Gliner et al26 achieved a rate of 83%. Ventricular fibrillation was successfully terminated by fewer than 3 shocks in 97% of patients.26 Animal studies suggest that these lower energies result in improved postshock myocardial function.49-50 Such data have resulted in biphasic waveforms becoming the most common waveform offered in AEDs, and the manufacturers of standard external defibrillators have begun to market biphasic waveform devices as well.
|
|
|
|
Figure 5. Defibrillation Waveforms
A, Monophasic truncated exponential (MTE). Current delivery begins at a peak and declines until the selected energy has been delivered, at which point energy delivery ceases. B, Biphasic truncated exponential (BTE). As with MTE, energy delivery begins at a peak. However, direction of current flow (polarity) reverses at a predetermined point and continues in this direction until the selected energy has been delivered or selected duration has been completed.
|
|
|
Nonrechargeable lithium-based batteries that can last up to 5 years without requiring service are rapidly replacing larger lead and nickel cadmium batteries as the energy source for the AED. At present, most devices automatically perform self-tests on a daily or weekly basis, alerting users when service is required. Such innovations in energy delivery and storage have led to marked reductions in maintenance requirements and reductions in both the size and cost ($3000-$4500 each) of AEDs (Table 1).
|
|
|
|
Table 1. Automated External Defibrillators
|
|
|
Use of the AED Within the Traditional EMS System
Several studies have demonstrated that EMTs and paramedics can safely and effectively use manual external defibrillators.51-54 Subsequent studies of AED use by these trained personnel demonstrated that the AED was equally safe and effective and suggested a possible survival advantage over use with the manual external defibrillator (Table 2).17-18,55 These data, along with improved portability and ease of use of the AED, have led to AEDs becoming standard equipment in many EMSs. Several communities have documented improved survival with the addition of EMT defibrillation using the AED,56-59 while 2 meta-analyses have demonstrated that defibrillation by basic life support providers reduced the relative risk of death for persons experiencing out-of-hospital cardiac arrest who are in ventricular fibrillation.60-61
|
|
|
|
Table 2. Comparison of Survival Rates From Out-of-Hospital Cardiac Arrest After Defibrillation With an Automated External Defibrillator (AED) vs a Manual Defibrillator*
|
|
|
In an effort to further reduce time to defibrillation and in response to evidence demonstrating the safety and ease of use of the AED, many communities expanded the role of defibrillation to trained first-responders (eg, police officers and firefighters) who often arrive at the scene of an arrest before paramedics. Studies of the use of the AED by such personnel have shown dramatically reduced time to defibrillation and enhanced survival in select communities (Table 3).62-68 Weaver et al62 showed that firefighters could deliver defibrillation with an AED 5 minutes earlier than paramedics could with a standard defibrillator. A study by White et al63 found that survival to hospital discharge in Rochester, Minn, was increased from 26% to 58% when patients in ventricular fibrillation were defibrillated by police. Mossesso et al64 showed that police use of the AED in Allegheny County, Pennsylvania, decreased time to defibrillation and was an independent predictor of survival to hospital discharge.
|
|
|
|
Table 3. Comparison of First-Responder Defibrillation With Paramedic/EMT Defibrillation
|
|
|
Despite these impressive results, several studies indirectly point out the inability of early defibrillation alone to overcome other deficiencies in the chain of survival. Kellermann et al69 showed that the impact on survival of adding first responder defibrillation to a fast-response urban EMS system was small despite a reduction in time to defibrillation. Low rates of bystander CPR (12%) were felt likely to have contributed to poor survival rates in this study. Sweeney et al70 also showed that the addition of AEDs to their EMS system in North Carolina failed to improve survival rates. Bystander CPR was impressively frequent, but delays in calling for EMS support and long call-processing times were noted. A large study by Stiell et al,71 involving more than 19 different communities in Ontario in which survival rates had previously been low (2.5%), showed implementation of a rapid defibrillation program to be an effective and inexpensive approach to improving out-of-hospital cardiac arrest survival.71-72 Much of the benefit to survival in this study, however, came from patients unlikely to benefit from early defibrillationthose initially found with pulseless electrical activity and asystolesuggesting that shorter response times and early CPR were important determinants of improved survival.73 A meta-analysis by Nichol et al74 concluded that while early defibrillation had the greatest relative impact on survival, increased rates of bystander CPR and the presence of advanced cardiac life supportcapable EMSs are also important determinants of survival. In summary, while each component of the chain of survival remains critical to improved survival, the AED is a vital addition to an EMS and a powerful tool in the effort to reduce time to defibrillation and improve survival from sudden cardiac death.
Outside the Traditional EMS System
Because the majority of cardiac arrests occur at home, several studies have examined the use of AEDs by family members of high-risk patients.75-76 Although these studies demonstrated the feasibility of training laypersons (eg, family members) to use an AED, researchers had difficulty with patient recruitment and obtained disappointing results. There is mounting evidence for the efficacy of ICDs in patients at increased risk for sudden cardiac death.77-79 This has limited enthusiasm for the placement of AEDs in the home of high-risk patients and primarily limited the role of the AED in the home to high-risk patients who either refuse an ICD or have a contraindication to ICD placement. However, these studies used earlier-generation AEDs and, given the lower costs and ease of use of the current devices, further study with the newer technology is warranted.
In 1991, Quantas Airlines initiated a program using AEDs on overseas flights and at major terminals. Quantas documented 46 cardiac arrests in a 64-month period in which long-term survival from arrests was 26%, comparable with the most effective prehospital emergency services.80 In data from American Airlines over a 2-year period, the AED was used on 200 patients. Of those with ventricular fibrillation, 6 of 15 (40%) were shocked and survived to hospital discharge with full neurologic and functional recovery. Specificity and sensitivity for ventricular fibrillation were 100%.81 Based on these experiences, several US and international airlines have installed AEDs or are planning to do so.82-83
Valenzuela et al84 looked at use of the AED by security personnel in casinos, a setting in which a large population is closely monitored. Use of the AED resulted in survival to discharge in 53% (56/105) of patients whose initial rhythm was ventricular fibrillation. Mean time to first defibrillation was only 4.4(2.9) minutes while mean time for arrival of paramedics was 9.8(4.3) minutes.
Public Access Defibrillation
How widespread the availability of the AED should be is unknown and whether AEDs should be placed in shopping malls, conventions centers, and large office buildings is largely untested. Becker et al,85 however, retrospectively looked at the potential benefit of placing AEDs in the higher-incidence sites of cardiac arrest in Seattle and King County, Washington and estimated that 134 cardiac arrest patients would have been provided treatment, with 8 to 32 lives saved over 5 years. The National Heart, Lung, and Blood Institute and the AHA are jointly supporting a multicenter, controlled, prospective clinical trial of PAD that should answer some of these important questions.86-88
In the Hospital
The use of the AED is not limited to prehospital patient care. While cardiac arrest survival rates in the coronary care unit can be as high as 90%, survival outside of these units falls off dramatically. Several minutes can elapse before conventional in-hospital code teams first attempt defibrillation.89-90 It has been shown that noncritical care nurses can learn to use the AED and retain the knowledge and skill over time.91 The use of the AED has made it possible to reduce time to defibrillation in noncritical care locations and, as a result, in-hospital AED programs are increasingly being encouraged.92-96 The AED has been used intraoperatively in high-risk patients with success. Advantages of the AED in the operating room included rapid response to ventricular arrhythmia; safe, hands-free operation; and minimal disruption of the surgical procedure.97
Cost to the Health Care System
Although based on multiple assumptions of cost and improvements in survival, initial cost-effectiveness analyses have suggested that PAD and first-responder defibrillation are economical in comparison to other common treatments for life-threatening illnesses.98-99 Nichol et al99 estimated that implementation of PAD by laypersons in an urban EMS system was associated with a median cost of $44 000 per additional quality-adjusted life-year saved and that the same program for police use was associated with a median cost of $27 200 per additional quality-adjusted life-year saved, consistent with the cost of other common medical interventions (ie, <$50 000 per quality-adjusted life-year). These data are based on multiple assumptions, including the cost to implement a PAD program and the survival rate from cardiac arrest, and must be evaluated with caution. Prospective randomized trials are needed to better answer these questions. The AHA Guidelines 2000 for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care currently call for PAD programs in areas where response times of fewer than 5 minutes, from EMS call to countershock, cannot be reliably achieved, and in areas where there exists a reasonable probability of at least 1 AED use in 5 years.86
Barriers to PAD
Physicians and legislators were initially slow to accept the concept of PAD. Recently, however, both state and federal governments have taken a more active role in promoting efforts at reduction of sudden cardiac death. In 1997 the 106th Congress passed the Cardiac Arrest Survivor Act of 1997, amending the Public Health Service Act to establish at the National Heart, Lung, and Blood Institute a program regarding lifesaving interventions for individuals who experience cardiac arrest. The success of manufacturers in developing more advanced AEDs has reduced concerns over inappropriate shocks and potential harm of defibrillation by laypersons. These improvements have buoyed the efforts to pass Good Samaritan laws, and currently 45 states have passed legislation protecting laypersons who use an AED in good faith.100 On May 20, 2000, the president proposed an initiative directing the creation of criteria for the placement of AEDs in federal buildings and on all commercial airlines in an effort to save up to 20 000 lives each year.101 These initiatives at the state and federal levels are paving the way for more widespread access to defibrillation as legal barriers to PAD, both perceived and real, are slowly eliminated.
CONCLUSION
Sudden cardiac death remains a major public health issue. Animal and human data demonstrate that early defibrillation improves survival, and that reductions in time to defibrillation can increase survival following sudden cardiac arrest. However, there are limitations to how quickly the EMSs can respond in many communities, particularly in rural and urban centers. The AED represents a major advance in the effort to achieve early defibrillation and further improve survival following out-of-hospital sudden cardiac arrest. By responding to the challenge to develop an AED that is more accurate, lightweight, affordable, and easy to use, AED manufacturers have helped make public access to defibrillation feasible. With help from the state and federal governments, manufacturers have helped overcome many of the obstacles to AED implementation. Automated external defibrillators are quickly becoming an integral part of the EMS and their presence in the community is increasing at a rapid rate. Additional studies are needed to determine how widespread the deployment of these lifesaving devices should be, provide more data on the cost-effectiveness of PAD, and further define the role of the AED in children and infants.
AUTHOR INFORMATION
Author Contributions: Study concept and design: Marenco, Wang, Link, Estes.
Acquisition of data: Marenco, Homoud.
Analysis and interpretation of data: Marenco, Wang, Link, Homoud.
Drafting of the manuscript: Marenco, Wang.
Critical revision of the manuscript for important intellectual content: Marenco, Wang, Link, Homoud, Estes.
Statistical expertise: Marenco, Link, Homoud.
Administrative, technical, or material support: Marenco.
Study supervision: Marenco, Wang, Link, Homoud, Estes.
Financial Disclosure: Dr Estes has received an honorarium from Medtronic for a talk.
Corresponding Author and Reprints: N. A. Mark Estes III, MD, New England Cardiac Arrhythmia Center, Division of Cardiology, New England Medical Center, 750 Washington St, Boston, MA 02111 (e-mail: nestes{at}lifespan.org).
Author Affiliations: New England Cardiac Arrhythmia Center, Division of Cardiology, Department of Medicine, New England Medical Center, Boston, Mass.
REFERENCES
 |  |
1. Gillum RF. Sudden coronary death in the United States: 1980-1985. Circulation. 1989;79:756-765.
FREE FULL TEXT
2. Zipes DP, Wellens HJ. Sudden cardiac death. Circulation. 1998;98:2334-2351.
FREE FULL TEXT
3. National Center for Health Statisics. Advance report of final mortality statistics, 1995. Monthly Vital Stat Rep. 1997;45:S2.
4. Becker LB, Ostrander MP, Barrett J, et al. Outcome of CPR in a large metropolitan area: where are the survivors? Ann Emerg Med. 1991;20:355-361.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
5. Lombardi G, Gallagher J, Gennis P. Outcome of out-of-hospital cardiac arrest in New York City: the prehospital arrest survival evaluation study. JAMA. 1994;271:678-683.
FREE FULL TEXT
6. Myerburg RJ, Kessler KM, Castellanos A. Sudden cardiac death: structure, function, and time-dependence of risk. Circulation. 1992;85(suppl 1):I2-I10.
7. Eisenberg MS, Bergner L, Hallstrom AP, Cummins RO. Sudden cardiac death. Sci Am. 1986;254:37-43.
WEB OF SCIENCE
| PUBMED
8. Cummins RO, Ornato JP, Thies WH, Pepe PE. Improving survival from sudden cardiac arrest: the chain of survival concept. Circulation. 1991;83:1832-1847.
FREE FULL TEXT
9. Weaver WD, Cobb LA, Hallstrom AP. Factors influencing survival after out-of-hospital cardiac arrest. J Am Coll Cardiol. 1986;7:752-775.
ABSTRACT
10. Nichol G, Stiell IG, Laupacis A, et al. A cumulative meta-analysis of the effectiveness of defibrillator-capable emergency medical services for victims of out-of-hospital cardiac arrest. Ann Emerg Med. 1999;34:517-525.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
11. Eisenberg MS, Hallstrom A, Bergner L. The ACLS score: predicting survival from out-of-hospital cardiac arrest. JAMA. 1981;246:50-52.
FREE FULL TEXT
12. Kouwenhoven WB. The development of the defibrillator. Ann Intern Med. 1969;71:449-458.
FREE FULL TEXT
13. Haskell WL. Cardiovascular complications during exercise training of cardiac patients. Circulation. 1978;57:920-924.
FREE FULL TEXT
14. Hossack KF, Hartwig R. Cardiac arrest asociated with supervised cardiac rehabilitation. J Card Rehabil. 1982;2:402-408.
15. Van Camp SP, Peterson RA. Cardiovascular complications of outpatient cardiac rehabilitation programs. JAMA. 1986;256:1160-1163.
FREE FULL TEXT
16. Cobb LA, Fahrenbruch CE, Walsh TR, et al. Influence of cardiopulmonary resuscitation prior to defibrillation in patients with out-of-hospital ventricular fibrillation. JAMA. 1999;281:1182-1188.
FREE FULL TEXT
17. Weaver WD, Hill D, Fahrenbuch CE, et al. Use of automated external defibrillators in the management of out-of-hospital cardiac arrest. N Engl J Med. 1988;319:661-666.
ABSTRACT
18. Stults KR, Brown DD, Kerber RE. Efficacy of an automated external defibrillator in the management of out-of-hospital cardiac arrest: validation of the diagnostic algorithm and the clinical experience in a rural environment. Circulation. 1986;73:701-709.
FREE FULL TEXT
19. Becker LB, Han BH, Meyer PM, et al. Racial differences in the incidence of cardiac arrest and subsequent survival. N Engl J Med. 1993;329:600-606.
FREE FULL TEXT
20. Weisfeldt ML, Kerber RE, McGoldrick RP, et al for the American Heart Association Task Force on Automatic External Defibrillation. Public access to defibrillation. Circulation. 1995;92:2763.
FREE FULL TEXT
21. Nichol G, Hallstrom AP, Kerber R, et al. American Heart Association Report on the Second Public Access Defibrillation Conference, April 17-19, 1997. Circulation. 1998;97:1309-1314.
FREE FULL TEXT
22. Diack AW, Welborn WS, Rullman RG. Use of an automatic cardiac resuscitator for emergency treatment of arrest. Med Instrum. 1979;13:78-81.
WEB OF SCIENCE
| PUBMED
23. Kerber RE, Becker LB, Bourland JD, et al. Automatic external defibrillators for public access defibrillation: recommendation for specifying and reporting arrhythmia analysis algorithm performance, incorporating new waveforms, and enhancing safety. Circulation. 1997;95:1677-1682.
FREE FULL TEXT
24. Cummins RO, Eisenberg M, Bergner L, Murray JA. Sensitivity, accuracy, and safety of an automatic external defibrillator. Lancet. 1984;2:318-320.
WEB OF SCIENCE
| PUBMED
25. Gundry JW, Comess KA, DeRook FA, Jorgenson D, Bardy GH. Comparison of naive sixth-grade children with trained professionals in the use of an automated external defibrillator. Circulation. 1999;100:1703-1707.
FREE FULL TEXT
26. Gliner BE, Jorgenson DB, Poole JE, et al for the LIFE Investigators. Treatment of out-of hospital cardiac arrest with low-energy impedance-compensating biphasic waveform automatic external defibrillators. Biomed Instrum Technol. 1998;32:631-644.
PUBMED
27. Stults KR, Brown DD, Cooley F, et al. Self-adhesive monitor/defibrillator pads improve pre-hospital defibrillation success. Ann Emerg Med. 1987;16:872-877.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
28. Carlson MD, Freeman CS, Garan H, Ruskin JN. Sensitivity of an automatic external defibrillator for ventricular tachyarrhythmias in patients undergoing electrophysiologic studies. Am J Cardiol. 1988;61:787-790.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
29. Poole JE, White RD, Kanz KG, et al. Low-energy impedance-compensating biphasic waveforms terminate ventricular fibrillation at high rates in victims of out-of-hospital cardiac arrest. J Cardiovasc Electrophysiol. 1997;8:1373-1385.
WEB OF SCIENCE
| PUBMED
30. White RD, Gliner BE. Transthoracic impedance does not affect defibrillation efficacy in cardiac arrest victims when shocks are delivered with an impedance-compensating biphasic waveform. Circulation. 1997;96(suppl 1):I-561.
31. Dalzell GWN, Adgey AAJ. Determinants of successful transthoracic defibrillation. Br Heart J. 1991;65:311-316.
FREE FULL TEXT
32. Kerber RE, Martins JB, Kienzle MG, et al. Energy, current, and success in defibrillation and cardioversion: clinical studies using an automated impedance-based method of energy adjustment. Circulation. 1988;77:1038-1046.
FREE FULL TEXT
33. Kerber RE. Electrical treatment of cardiac arrythmias: defibrillation and cardioversion. Ann Emerg Med. 1993;22:296-301.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
34. Dalzell GWN, Cunningham SR, Anderson J, Adgey AA. Electrode pad size, transthoracic impedance and success of external ventricular defibrillation. Am J Cardiol. 1989;64:741-744.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
35. Hoyt R, Grayzel J, Kerber RE. Determinants of intracardiac current in defibrillation: experimental studies in dogs. Circulation. 1981;64:818-823.
FREE FULL TEXT
36. Bardy GH, Marchlinski FE, Sharma AD, et al. Multicenter comparison of truncated biphasic shocks and standard damped sine wave monophasic shocks for transthoracic ventricular defibrillation. Circulation. 1996;94:2507-2514.
FREE FULL TEXT
37. Bardy GH, Gliner BE, Kudenchuk PJ, et al. Truncated biphasic pulses for transthoracic defibrillation. Circulation. 1995;91:1768-1774.
FREE FULL TEXT
38. Walcott GP, Melnick SB, Chapman FW, et al. Relative efficacy of monophasic and biphasic waveforms for transthoracic defibrillation after short and long durations of ventricular fibrillation. Circulation. 1998;98:2210-2215.
FREE FULL TEXT
39. Greene HL, DiMarco JP, Kudenchuk PJ, et al for the Biphasic Waveform Defibrillation Investigators. Comparison of monophasic and biphasic defibrillating pulse waveforms for transthoracic cardioversion. Am J Cardiol. 1995;75:1135-1139.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
40. Wilson CM, Bailey A, Allen JD, Anderson J, Adgey AA. Transthoracic defibrillation threshold of sine and trapezoidal waveforms in defibrillation. J Electrocardiol. 1989;22:241-247.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
41. Swartz JF, Fletcher RF, Karasik PJ, et al. Optimization of biphasic waveforms for human nonthoracotomy defibrillation. Circulation. 1993;88:2646-2654.
FREE FULL TEXT
42. Bardy GH, Gliner BE, White RD. Comparison of human defibrillation efficacy rates in short duration and long duration ventricular fibrillation using monophasic and biphasic transthoracic shocks. Circulation. 1997;96(suppl):I-560.
43. Kerber RE, Spencer KT, Kallok MJ, et al. Overlapping sequential pulses: a new waveform for transthoracic defibrillation. Circulation. 1994;89:2369-2379.
FREE FULL TEXT
44. Peleska B. Optimal parameters of electrical impulses for defibrillation by condenser discharges. Circ Res. 1966;18:10-17.
FREE FULL TEXT
45. Yamanouchi Y, Brewer JE, Olson KF, et al. Fully discharging phases: a new approach to biphasic waveforms for external defibrillation. Circulation. 1999;100:826-831.
FREE FULL TEXT
46. Yamanouchi Y, Brewer JE, Mowrey KA, Donohoo AM, Wilkoff BL, Tchou P. The optimal small capacitor biphasic waveform for external defibrillation: influence of phase-1 tilt and phase-2 voltage. Circulation. 1998;98:2487-2493.
FREE FULL TEXT
47. Cummins RO, Hazinski F, Kerber RE, et al for the American Heart Association Committee on Emergency Cardiovascular Care. Low-energy biphasic waveform defibrillation: evidence-based review applied to emergency cardiac care guidelines. Circulation. 1998;97:1654-1667.
FREE FULL TEXT
48. Ideker RE, Charbonnier F. Established and alternative waveforms. Workshop report presented at: Second American Heart Association Public Access Defibrillation Conference; April 19, 1997; Washington, DC.
49. Tang W, Weil MH, Sun S, et al. Low-energy biphasic waveform defibrillation reduces the severity of postresuscitation myocardial dysfunction after prolonged arrest. Crit Care Med. 2000;28:N222-N224.
50. Tang W, Weil MH, Sun S, et al. The effects of biphasic and conventional monophasic defibrillation on postresuscitation myocardial function. J Am Coll Cardiol. 1999;34:815-822.
FREE FULL TEXT
51. Bachman JW, McDonald GS, O'Brien PC. A study of out-of-hospital cardiac arrests in northeastern Minnesota. JAMA. 1986;256:477-483.
FREE FULL TEXT
52. Olson DW, LaRochelle J, Fark D, et al. EMT-defibrillation: the Wisconsin experience. Ann Emerg Med. 1989;18:806-811.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
53. Eisenberg MS, Copass MK, Hallstrom AP. Treatment of out-of hospital cardiac arrest with rapid defibrillation by emergency medical technicians. N Engl J Med. 1980;302:1379-1383.
ABSTRACT
54. Stults KR, Brown DD, Schug VL. Prehospital defibrillation performed by emergency medical technicians in rural communities. N Engl J Med. 1984;310:219-223.
ABSTRACT
55. Cummins RO, Eisenberg MS, Litwin PE. Automatic external defibrillators used by emergency medical technicians: a controlled clinical trial. JAMA. 1987;257:1605-1610.
FREE FULL TEXT
56. Gray AJ, Redmond AD, Martin MA. Use of the automatic external defibrillator-pacemaker by ambulance personnel: the Stockport experience. Br Med J (Clin Res Ed). 1987;294:1133-1135.
FREE FULL TEXT
57. Jakobsson J, Nyquist O, Rehnquist N. Effects of early defibrillation of out-of-hospital cardiac arrest patients by ambulance personnel. Eur Heart J. 1987;8:1189-1194.
FREE FULL TEXT
58. Eisenberg MS, Hallstrom AP, Copass MK. Treatment of ventricular fibrillation: emergency medical technician defibrillation and paramedic services. JAMA. 1984;251:1723-1726.
FREE FULL TEXT
59. Vukov IF, White RD, Backman JW. New perspectives on rural EMT defibrillation. Ann Emerg Med. 1988;17:318-321.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
60. Watts DD. Defibrillation by basic emergency medical technicians: effect on survival. Ann Emerg Med. 1995;26:635-639.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
61. Auble TE, Menegazzi JL, Paris PM. Effect of out-of-hospital defibrillation by basic life support providers on cardiac arrest mortality: a meta-analysis. Ann Emerg Med. 1995;25:642-648.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
62. Weaver WD, Copass MK, Bufi D, Ray R, Hallstrom AP, Cobb LA. Improved neurologic recovery and survival after early defibrillation. Circulation. 1984;69:943-948.
FREE FULL TEXT
63. White RD, Vukov LF, Bugliosi TF. Early defibrillation by police: initial experience with measurement of critical time intervals and patient outcome. Ann Emerg Med. 1994;23:1009-1013.
WEB OF SCIENCE
| PUBMED
64. Mossesso VN Jr, Davis EA, Auble TE, Paris PM, Yealy DM. Use of automated external defibrillators by police officers for treatment of out-of-hospital cardiac arrest. Ann Emerg Med. 1998;32:200-207.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
65. White RD, Asplin BR, Bugliosi TF, Hankins DG. High discharge survival rate after out-of-hospital ventricular fibrillation with rapid defibrillation by police and paramedics. Ann Emerg Med. 1996;28:480-485.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
66. Shuster M, Keller JL. Effect of fire department first-responder automated defibrillation. Ann Emerg Med. 1993;22:721-727.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
67. Weaver WD, Copass MK, Hill D, Fahrenbruch C, Hallstrom AP, Cobb L. Cardiac arrest treated with a new automatic external defibrillator by out-of-hospital first responders. Am J Cardiol. 1986;57:1017-1021.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
68. Davis EA, Mosesso VN Jr. Performance of police first-responders in utilizing automated external defibrillation on victims of sudden cardiac arrest. Prehosp Emerg Care. 1998;2:101-107.
PUBMED
69. Kellermann AL, Hackman BB, Somes G, Kreth TK, Nail L, Dobyns P. Impact of first-responder defibrillation in an urban emergency medical services system. JAMA. 1993;270:1708-1713.
FREE FULL TEXT
70. Sweeney TA, Runge JW, Gibbs JM, Schafermeyer RW, Norton HJ, Boyle-Whitesel MJ. EMT defibrillation does not increase survival from sudden cardiac death in a two-tiered urban-suburban EMS system. Ann Emerg Med. 1998;31:234-240.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
71. Stiell IG, Wells GA, Field BJ, et al. Improved out-of-hospital cardiac arrest survival through the inexpensive optimization of an existing defibrillation program. JAMA. 1999;281:1175-1181.
FREE FULL TEXT
72. Brison RJ, Davidson JR, Dreyer JF, et al. Cardiac arrest in Ontario: circumstances, community reponse, role of prehospital defibrillation and predictors of survival. CMAJ. 1992;147:191-199.
ABSTRACT
73. Asplin BR, Yealy DM, White RD. Improving survival following out-of-hospital cardiac arrest. JAMA. 1999;282:1033-1034.
FREE FULL TEXT
74. Nichol G, Stiell IG, Laupacis A, et al. A cumulative meta-analysis of the effectiveness of defibrillator-capable emergency medical services for victims of out-of-hospital cardiac arrest. Ann Emerg Med. 1999;34:517-525.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
75. McDaniel CM, Berry VA, Haines DE, et al. Automatic external defibrillation of patients after myocardial infarction by family members: practical aspects and psychological impact of training. Pacing Clin Electrophysiol. 1988;11:2029-2034.
FULL TEXT
| PUBMED
76. Eisenberg MS, Moore J, Cummins RO, et al. Use of the automatic external defibrillator in homes of survivors of out-of-hospital ventricular fibrillation. Am J Cardiol. 1989;63:443-446.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
77. The Antiarrhythmics Versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmic-only therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrythmias. N Engl J Med. 1997;337:1576-1583.
FREE FULL TEXT
78. Siebels J, Kuck KH for the CASH investigators. Implantable cardioverter defibrillator compared with antiarrhythmic drug treatment in cardiac arrest survivors (the Cardiac Arrest Study Hamburg). Am Heart J. 1994;127:1139-1144.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
79. Moss AJ, Hall J, Cannom DS, et al. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia: the Multi-Center Automatic Defibrillator Implantation Trial. N Engl J Med. 1996;335:1933-1940.
FREE FULL TEXT
80. O'Rourke MF, Donaldson EE, Geddes JS. An airline cardiac arrest program. Circulation. 1997;96:2849-2853.
FREE FULL TEXT
81. Page RL, Joglar JA, Kowal RC, et al. Use of automated external defibrillators by a U.S. airline. N Engl J Med. 2000;343:1210-1216.
FREE FULL TEXT
82. Zimmermann PG, Campbell L. Automatic external defibrillators on commercial airplanes. J Emerg Nurs. 1997;23:340-342.
FULL TEXT
| PUBMED
83. Smith SC, Hamburg RS. Automated external defibrillators: time for federal and state advocacy and broader utilization. Circulation. 1998;97:1321-1324.
FREE FULL TEXT
84. Valenzuela TD, Roe DJ, Nichol G, Clark LL, Spaite DW, Hardman RG. Outcomes of rapid defibrillation by security officers after cardiac arrest in casinos. N Engl J Med. 2000;343:1206-1209.
FREE FULL TEXT
85. Becker L, Eisenberg M, Fahrenbruch C, Cobb L. Public locations of cardiac arrest: implications for public access defibrillation. Circulation. 1998;97:2106-2109.
FREE FULL TEXT
86. The American Heart Association in Collaboration with the International Liaison Committee on Resuscitation. Guidelines 2000 for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2000;102(suppl 1):I60-I76.
87. Robertson RM. Sudden death from cardiac arrest: improving the odds. N Engl J Med. 2000;343:1259-1260.
FREE FULL TEXT
88. Hallstrom AP. Public Access Defibrillation (PAD) Community Trial [abstract] [CRISP database online]. Grant 1N01HC95177. Available at: http://commons.cit.nih.gov/crisp/CRISP.generate_ticket. Access confirmed.
89. Lazzam C, McCans J. Predictors of survival of in-hospital cardiac arrest. Can J Cardiol. 1991;7:113-116.
PUBMED
90. Dickey W, Adgey A. Mortality within hospital from ventricular fibrillation outside the hospital. Br Heart J. 1992;67:334-338.
FREE FULL TEXT
91. Mancini ME, Kaye W. In-hospital first-responder automated external defibrillation: what critical care practitioners need to know. Am J Crit Care. 1998;7:314-319.
ABSTRACT
92. Kaye W, Mancini ME, Guiliano KK, Richards N. Strengthening the in-hospital chain of survival with rapid defibrillation by first responders using automated external defibrillators: training and retention issues. Ann Emerg Med. 1995;25:163-168.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
93. Kaye W, Mancini ME, Richards N. Organizing and implementing a hospital-wide first-responder automated external defibrillation program: strengthening the in-hospital chain of survival. Resuscitation. 1995;30:151-156.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
94. Kaye W, Mancini ME. Improving outcome from cardiac arrest in the hospital with a reorganized and strengthened chain of survival: an American view. Resuscitation. 1996;31:181-186.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
95. Destro A, Marzaloni M, Sermasi S, Rossi F. Automated external defibrillators in the hospital as well? Resuscitation. 1996;31:39-43.
PUBMED
96. Spearpoint KG, McLean CP, Zideman DA. Early defibrillation and the chain of survival in "in-hospital" adult cardiac arrest: minutes count. Resuscitation. 2000;44:165-169.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
97. Jones FD, Fahy BG. Intraoperative use of automated external defibrillator. J Clin Anesth. 1999;11:336-338.
FULL TEXT
|
WEB OF SCIENCE
| PUBMED
98. Jermyn BD. Cost-effectiveness analysis of a rural/urban first-responder defibrillation program. Prehosp Emerg Care. 2000;4:43-47.
PUBMED
99. Nichol G, Hallstrom AP, Ornato JP, et al. Potential cost-effectiveness of public access defibrillation in the United States. Circulation. 1998;97:1315-1320.
FREE FULL TEXT
100. SoRelle R. States set to pass laws limiting liability for lay users of automated defibrillators. Circulation. 1999;99:2606-2607.
FREE FULL TEXT
101. Clinton William Jefferson. White House press release. June 1, 2000.
Clinical Cardiology Section Editor: Michael S. Lauer, MD, Contributing Editor, JAMA.
CiteULike Connotea Del.icio.us Digg Reddit Technorati Twitter
What's this?
RELATED LETTER
Informed Consent for Public Automated External Defibrillation
Anne Scheetz
JAMA. 2001;286(1):47.
EXTRACT
| FULL TEXT
RELATED ARTICLES
March 7, 2001
JAMA. 2001;285(9):1229-1230.
EXTRACT
| FULL TEXT
Heart Emergencies Caused by Irregular Heartbeat
JAMA. 2001;285(9):1248.
PDF
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
 |
Modified alternating current defibrillation: a new defibrillation technique
Rosenheck et al.
Europace 2009;11:239-244.
ABSTRACT
| FULL TEXT
Can Home AEDs Improve Survival?
Callans
NEJM 2008;358:1853-1855.
FULL TEXT
Prediction and Prevention of Sudden Cardiac Arrest: Lessons Learned in Schools
Estes
Circulation 2007;116:1341-1343.
FULL TEXT
ACC/AHA/ESC 2006 Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death--Executive Summary: A Report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death)
Developed in Collaboration With the European Heart et al.
J Am Coll Cardiol 2006;48:1064-1108.
FULL TEXT
ACC/AHA/ESC 2006 Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death: A Report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death)
Developed in Collaboration With the European Heart et al.
J Am Coll Cardiol 2006;48:e247-e346.
FULL TEXT
ACC/AHA/ESC 2006 guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death--executive summary: A report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients with Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death) Developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society.
Zipes et al.
Eur Heart J 2006;27:2099-2140.
FULL TEXT
ACC/AHA/ESC 2006 guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: A report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death) Developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society
Writing Committee Members et al.
Europace 2006;8:746-837.
FULL TEXT
Recalls and safety alerts affecting automated external defibrillators.
Shah and Maisel
JAMA 2006;296:655-660.
ABSTRACT
| FULL TEXT
Automated external defibrillators--device reliability and clinical benefits.
Estes
JAMA 2006;296:700-702.
FULL TEXT
The Automated External Defibrillator: Clinical Benefits and Legal Liability
England et al.
JAMA 2006;295:687-690.
FULL TEXT
A 2-Tiered Approach to In-Hospital Defibrillation: Nurses Respond to a Trial of Using Automated External Defibrillators as Part of a Code-Team Protocol
Kyller and Johnstone
Crit Care Nurse 2005;25:25-33.
FULL TEXT
Task Force 10: Automated external defibrillators
Myerburg et al.
J Am Coll Cardiol 2005;45:1369-1371.
FULL TEXT
Automatic external defibrillation in a 6 year old
Konig et al.
Arch. Dis. Child. 2005;90:310-311.
ABSTRACT
| FULL TEXT
A program encouraging early defibrillation results in improved in-hospital resuscitation efficacy
Zafari et al.
J Am Coll Cardiol 2004;44:846-852.
ABSTRACT
| FULL TEXT
Temporal Trends in Coronary Heart Disease Mortality and Sudden Cardiac Death From 1950 to 1999: The Framingham Heart Study
Fox et al.
Circulation 2004;110:522-527.
ABSTRACT
| FULL TEXT
Defibrillators
Journal of the American Dental Association 2004;135:366-367.
FULL TEXT
Cardiac Arrest Outside of a Hospital: How Can We Improve Results of Resuscitation?
Wellens et al.
Circulation 2003;107:1948-1950.
FULL TEXT
Impact of Community-Wide Police Car Deployment of Automated External Defibrillators on Survival From Out-of-Hospital Cardiac Arrest
Myerburg et al.
Circulation 2002;106:1058-1064.
ABSTRACT
| FULL TEXT
AEDS at School: What are the Issues?
Jackson
NASN School Nurse 2002;17:17-18.
The PULSE Initiative: Scientific Priorities and Strategic Planning for Resuscitation Research and Life Saving Therapies
Becker et al.
Circulation 2002;105:2562-2570.
FULL TEXT
Automated External Defibrillators in Health/Fitness Facilities: Supplement to the AHA/ACSM Recommendations for Cardiovascular Screening, Staffing, and Emergency Policies at Health/Fitness Facilities
Balady et al.
Circulation 2002;105:1147-1150.
FULL TEXT
Bringing order from chaos: The challenge of training lay responders to use automated external defibrillators and provide cardiopulmonary resuscitation
Woollard
Health Education Journal 2002;61:197-211.
ABSTRACT
Implantable Cardioverter Defibrillator: A Volkswagon or a Rolls Royce: How Much Will You Pay to Save a Life?
Hallstrom and Zipes
Circulation 2001;104
:e148-e148.
FULL TEXT
Saving Time Saves Lives
Zipes
Circulation 2001;104:2506-2508.
FULL TEXT
Cost-effectiveness of Automated External Defibrillators on Airlines
Groeneveld et al.
JAMA 2001;286:1482-1489.
ABSTRACT
| FULL TEXT
Primary Prevention of Sudden Death With Implantable Defibrillator Therapy in Patients With Cardiac Disease: Can We Afford to Do It? (Can We Afford Not To?)
Exner et al.
Circulation 2001;104:1564-1570.
FULL TEXT
Informed Consent for Public Automated External Defibrillation
Scheetz
JAMA 2001;286:47-47.
FULL TEXT
|