False-Positive Direct Fluorescent Antibody Testing for Legionella
LCDR David A. Johnson, MC
USNR
CAPT Kenneth F. Wagner, MC
USN
LCDR James Blanks, MSC
USN
JACQUELYN SLATER, MS
Portsmouth Naval Hospital Portsmouth, Va
JAMA. 1985;253(1):40-41.
 |
 |
| Since this article does not have an abstract, we have provided the first 150 words of the full text PDF and any section headings. |
|
 |
 |
To the Editor.—
Direct fluorescent antibody (DFA) testing of lung tissue or respiratory tract secretions has proved to be an expedient, reliable means for establishing the diagnosis of legionnaires' disease. The specificity of the DFA test on sputum is about 95% using the polyvalent reagent supplied by the Centers for Disease Control (CDC) in Atlanta.1 False-positives with DFA have been described in cases wherein filter water and/or reagents for rinsing contain Legionella.2 False-positive cross-reactions have been reported with isolates of Pseudomonas, Flavobacterium- xanthomonas group, and Bacteroides fragilis.3 We have observed two separate instances when Hemophilus influenzae in the sputum cross-reacted with pool C of the polyvalent assay (Legionella bozemanii, Legionella micadadei, Legionella gormanii, and Legionella longbeachae), resulting in 4+ fluorescence ( 5 bacilli per high-power field, magnification x 1,000) consistent with the CDC criteria for a positive test result. Sputum for culture and monovalent DFA test results were